Exercise Testing System, Model ASA26
- Ask for details on 0724 175 996
Model: ASA26
DESCRIPTION:
1. BS XL DIAGNOSTIC SYSTEM
12-channel recording, according to Einthoven, Goldberg, Wilson and Graber
Automatic measurement of HR, ST level and slope
Can interact with a treadmill or exercise bike
Option to re-analyse completed examinations
Report printing and archiving
Recording and observation
Simultaneous recording and observation of the ECG signal of all 12 leads on the monitor
Recording and observation of the average P-QRS-T complexes from all 12 leads, ST level and slope in online mode
Observation of the measured points – BL, J and ST, with the option of manual correction
Recording and observation of the current HR
Presentation of the examined phase parameters
PMHR control, observation of % PMHR
Recording and observation of MET, DPR
Review of the trends of all parameters during the examination
Automatic arrhythmia analysis
Automatic blood pressure measurement option
Alarms for monitored parameters and arrhythmias
Protocols:
Pre-programmed protocols
Predefined protocols
RAMP protocol
Printing and archiving:
Printing on a laser printer in A4 format
Real-time printing of ECG curves
Printing and archiving of examination reports, which contain:
Patient data
Full disclosure of the entire ECG recording
Strip selection
Average P-QRS-T complexes
Summary table
Load value, HR
ST level and slope
MET
Option to review reports on screen before printing
Sensitivity: 2.5, 5, 10, 20 mm/mV
Speed: 25, 50, 100 mm/s
Filters: baseline filters
Mains interference filter: 50 Hz
Muscle interference filter: 25 Hz, 35 Hz
Certificate: CE 0197, EN ISO 9001, EN ISO 13485, MDD 93/42/EEC
Technical data:
Supply voltage: 230V, 50 Hz
Maximum power consumption: < 500 W
Communication interface: LAN, 6 x USB, 2 x RS232
Peripheral devices: DVD writer, laser printer, 24″ LCD monitor
Electrical protection: Class 1
Weight: 70 kg
Dimensions: 71 W x 78 D x 91 H (135 H with monitor) cm
2. as500 ECG SIGNAL TRANSMITTER
ECG signal: standard 12 channels, according to Einthoven, Goldberg, Wilson and Graber
Operating mode, manual and automatic
Printing via:
3/6/12 channels
A4 printing via an external laser printer
Pacemaker detection
INOP detection
Sensitivity: 2.5/5/10/20 mm/mV
Speed: 5/10/25/50 mm/s
Communication interface: wireless communication with the PC via radio transmission (2.4 MHz) up to 10 m
Weight: 250 g
Dimensions: 130 L x 96 W x 30 H mm
3. AS B612 TREADMILL
Incline: 0-25%, selectable in 0.5% steps
Inclinometer: provides automatic inclination and adjustment of the treadmill
Zero Start function
Safety stop buttons (2), one on the handrails and one for medical staff
The ECG module includes:
ECG signals, with 12 leads in diagnostic mode
HR measuring range: 25 – 300 bpm ± 2%
HR resolution: 1 bpm
ST wave change monitoring range: ± 3 mm, ± 0.2 mm at 0.1 mm resolution for 10mm/mV
Sensitivity: 2.5/5/10/20mm/mV
Recording speed: 25/50/100 mm/s
Filter:
Mains interference: none, 50 Hz, 60 Hz
Muscle interference: none, 25 Hz, 35 Hz, 45 Hz
Baseline filter: none, 0.15, 0.45, 0.75, 1.5 Hz
Connection interface: RS-232 and USB – connection to the PC, which controls the treadmill operation
Standard accessories:
Patient cable: KEKG 51
Disposable electrodes: LLL-510
Abrasive paste EVERY
Technical data:
Supply voltage: 230 V, 50 Hz
Maximum load capacity: 180 kg
Treadmill weight: 185 kg
Applicable standards: PN-EN 60601-1, PN-EN 957-1, PN-EN 957-6
Electromagnetic compatibility: PN-EN 60601-1-2
Medical equipment class: IIa
Electrical protection against electric shock (PN-EN 60601-1): Class SA
Applied parts (PN-EN 60601-1): Type CF, defibrillation-proof
IP protection class: IP X0
Certificate: CE0197, EN ISO 9001, EN ISO 13485, MDD 93/42/EEC
Running belt size: 1,500 L x 500 W mm
Overall size: 1999 L x 747 W x 1217 H mm
4. AS CRG200 BICYCLE ERGOMETER
Braking: controlled brake with torque measurement of the load level
Load level: 25-400 W +/- 5 W at a power level of 50W, ±10% above 50 W
Microprocessor control
The load does not depend on speed
Rotation speed 20-130 rpm +/- 4 rpm
Functional characteristics
Integrated ECG module for recording 12 standard leads, in accordance with Einthoven, Goldberg, Wilson and Cabrer
Automatic measurement of HR and ST level
Automatic ergometer control
Signals when the speed needs to be changed
Speed adjustable to the patient’s individual needs
Suitable for examining paediatric patients taller than 140 cm
Adjustable saddle and handlebar
Protection against defibrillator pulse
INOP detection
Integrated ECG module
ECG signals: 12 standard leads in diagnostic mode
HR measuring range: 15-240 bpm
HR measurement error: +/- 2%
HR resolution: 1 bpm
ST segment monitoring range: +/- 3mm (10mm / mV)
ST measurement error: +/- 0.2mm (10mm / mV)
ST resolution: 0.1mm (10mm / mV)
Consultancy
We offer personalised consultancy and support for medical projects, from equipment to integrated solutions.
Nationwide delivery
We deliver nationwide, directly to the customer's location, or you can collect from our distribution points.
SEAP public tenders
On request, we can list the products you need on the SEAP platform for public procurement.
Related products